66 7 Alarms
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
Low PIP
Patient alarms
Leak or
disconnection.
Check patient,
Check patient
connection.
Check tubing
system for leaks.
Check setting of
Pinsp.
II
8
Charge
battery
(<60min)
Electrical
power
Ventilator is not
connected to
mains supply.
Power failure.
Internal battery
remaining charge
less than 60
minutes.
Switch to mains
supply.
Alarm can be silenced
with Alarm Silence
button. The device
continues in battery
mode without
interruption.
II
8
Charge
battery
(<30min)
Electrical
power
Ventilator is not
connected to
mains supply.
Remaining battery
time < 30 min.
Switch to mains
supply.
Alarm can be silenced
with Alarm Silence
button. The device
continues in battery
mode without
interruption.
II
8
High breath
rate
Patient alarms
Hyperventilation
self-trigger
Check patient,
Adjust frequency
Increase trigger
threshold
II
8
High tube leak
Patient alarms
Upper alarm limit
exceeded
Check patient,
Check tubing
system for leaks,
Adjust limit value.
II
8
High minute
volume
Patient alarms
Lung compliance
has increased.
Resistance has
decreased.
Hyperventilation.
Check patient,
Check ventilation
settings.
Adjust settings
II
8
Low minute
volume
Patient alarms
Lung compliance
has decreased.
Resistance has
increased.
Spontaneous
breathing has
stopped or is
declining
Check patient,
Check tubing
system for
obstructions,
Check ventilation
settings
II
8

Table of Contents

Questions and Answers: