7 Alarms 67
fabian +nCPAP evolution | SW V5.1.x
Ref: 122003.EN / Date : 26Jan2021
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
Low PEEP
Patient alarms
Leak or
disconnection.
Insp. flow or exp.
flow set too low.
Check patient,
Check tubing
system for tight
connection.
Increase flow.
II
8
VTe not
reached /
check settings
System limits
Inspiratory pressure
default is reached
prematurely.
Patient compliance
has worsened
Check patient,
Adjust inspiratory
pressure setting.
Increase
inspiratory period.
Adjust volume
setting
II
8
Power failure
Electrical
power
Power failure/
Ventilator was
unplugged from
mains power supply
and is operating on
battery power.
Switch to mains
supply.
Check power cord
for loose
connection.
Alarm can be
silenced with
Alarm Slience
button.
The device
continues in
battery mode
without
interruption.
III
9
Tidal Volume
limited
System limits
Upper alarm limit
exceeded
Adjust settings.
III
9