7351 Series Digital Multimeter Operation Manual
6.7.4 SCPI Command Reference
6-34
2. Measurement condition command
Item SCPI command
Description
Initial
value
Measure-
ment condi-
tion
Measurement
range
[:SENSe]:<function>:RANGe
:AUTO {ON,1|OFF,0} *1
Auto-range setting
ON, 1: Auto-range ON
OFF, 0: Auto-range OFF
This function cannot be set for the continuity, diode, and fre-
quency measurements.
[:SENSe]:<function>:RANGe
:AUTO? *1
Reply: 1 or 0
[:SENSe]:<function>:RANGe
{<real>|MAXimum|MINimum} *1
Fixed range setting
<real>:0
Setting value Maximum value for the measure-
ment range
(For the frequency measurement, the maximum value is
750(V))
This function cannot be set for the continuity and diode mea-
surements.
[:SENSe]:<function>:RANGe?
{MAXimum|MINimum} *1
Reply: +d.dddddE
dd
The reply is the maximum value for the specified range.
Sampling rate [:SENSe]:<function>:SRATe
{FAST|MED|SLOW|SSLow} *1
FAST: FAST
MED: MED
SLOW: SLOW1
SSLow: SLOW2
The setting is applied to all functions.
[:SENSe]:<function>:SRATe? *1 Reply: FAST, MED, SLOW, or SSL
Number of
display digits
[:SENSe]:<function>:DIGits
{3|4|5|MAXimum|MINimum} *1
3: 3 1/2-digit display
4: 4 1/2-digit display
5: 5 1/2-digit display
The setting is applied to all functions.
[:SENSe]:<function>:DIGits?
{MAXimum|MINimum} *1
Reply: 3, 4, or 5
Auto-zero [:SENSe]:ZERO:AUTO
{ON,1|OFF,0|ONCE}
ON, 1: ON
OFF, 0: OFF
ONCE: (Auto-zero is set to OFF after executing once.)
[:SENSe]:ZERO:AUTO? Reply: 1 or 0
Continuity
threshold con-
stant
[:SENSe]:CONTinuity:THReshold
{<real>|MAXimum|MINimum}
<real>: 1 to 1000 () setting resolution: 1 (10)
[:SENSe]:CONTinuity:THReshold?
{MAXimum|MINimum}
Reply: dddd
*1: <Function>: Specifies a function, which is not enclosed by double quotes, for "measurement function".