7351 Series Digital Multimeter Operation Manual
4.2.1 Measurement Functions
4-13
Figure 4-9 AC Current Measurement (ACI and ACI (AC+DC))
Measurement method TrueRMS
Input range 5% or more of a full scale
Crest factor 3 (This is restricted to the current protection fuse rating.)
Filter response time Approx. 1 sec (Time until the measurement value reaches within 0.1%
of the final value)
Auto-range restrictions
The 10 A range is a fixed range. The auto-range cannot function between the 10 A range and 200 mA,
2000 mA ranges.
4.2.1.6 Continuity Measurement
1. Connect a DUT to the and COM terminals on the front panel.
The continuity indicator ( ) is activated.
2. Press and .
The resistance is measured in 2W. If the measured resistance is lower than the continuity threshold
constant, the buzzer sounds.
Measurement range 2000
Measurement current 1 mA
Continuity threshold constant 1 to 1000
WARNING: Do not apply voltage that exceeds the maximum allowable applied voltage. Fire or electric
shocks due to the failure of this instrument may occur.
10 A or less
10 Arms or less
250 V or less
2 A or less
2 Arms or less