7351 Series Digital Multimeter Operation Manual
4.2.3 Changing the Measurement Speed and Display Digit
4-18
Table 4-11 Measurement Speed (b) when in the HOLD ON (Trigger source: EXT)
Sampling rate
Measurement speed (Measurement time)
Integration
time
DCV
DCI
2W
LP-2W
ACV
ACI
Continuity
Diode
ACV(AC+DC)
ACI(AC+DC)
FAST 80 times/s (12.5 ms) 45 times/s (22 ms) 2 ms
MED (50 Hz) 32 times/s (31 ms) 17 times/s (58 ms) 1 PLC
MED (60 Hz) 37 times/s (27 ms) 19 times/s (52 ms) 1 PLC
SLOW1 9 times/s (110 ms) 4.5 times/s (218 ms) 100 ms
SLOW2 4.7 times/s (210 ms) 2.3 times/s (418 ms) 200 ms
Conditions: Auto-range: OFF, Auto-zero: OFF, Calculation: OFF, Display: OFF, Trig-
ger delay: 0 sec, and GPIB data output
Table 4-12 Measurement Speed (b) when in the HOLD ON (Trigger source: BUS)
Sampling rate
Measurement speed (Measurement time)
Integration
time
DCV
DCI
2W
LP-2W
ACV
ACI
Continuity
Diode
ACV(AC+DC)
ACI(AC+DC)
FAST 74 times/s (13.5 ms) 42 times/s (23.5 ms) 2 ms
MED (50 Hz) 31 times/s (32 ms) 16 times/s (60 ms) 1 PLC
MED (60 Hz) 35 times/s (28 ms) 18 times/s (53 ms) 1 PLC
SLOW1 9 times/s (111 ms) 4.5 times/s (219 ms) 100 ms
SLOW2 4.7 times/s (211 ms) 2.3 time/s (419 ms) 200 ms
Conditions: Auto-range: OFF, Auto-zero: OFF, Calculation: OFF, Display: OFF, Trig-
ger delay: 0 sec, and GPIB