EasyManua.ls Logo

ADCMT 7351A - Auto-Zero Operation

ADCMT 7351A
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7351 Series Digital Multimeter Operation Manual
5.4 Auto-Zero Operation
5-10
5.4 Auto-Zero Operation
Auto-zero function eliminates the offset error in the measurement system of this instrument.
If Auto-zero is set to ON, the internal offset is measured and the offset error is eliminated because the offset
is subtracted from the input measurement value. However, the measurement time is doubled.
Set Auto-zero in ‘A.ZERO’ in in .
Auto-zero is performed at the current measurement range in the measurement function and range shown in
the following table.
Auto zero can be performed only once according to the measurement condition or measurement timing.
If ONCE is set and Auto-zero is performed, the offset error is eliminated and then Auto-zero is turned off.
Function Auto-zero
DCV Can be performed
ACV
AC+DCV
-
2W
LP-2W
Can be performed
DCI Can be performed
ACI
AC+DCI
-
FREQ -
Diode Can be performed
Continuity Can be performed
1 MEAS
MENU

Table of Contents

Related product manuals