7351 Series Digital Multimeter Operation Manual
7.3 General Specifications
7-15
7.3 General Specifications
Item Specifications
Operation environment Ambient temperature:0C to +50C
Relative humidity: 85%RH or less no condensation
Storage environment Ambient temperature:-25C to +70C
Warm-up time 60 minutes or more
Display 7-digit fluorescent display A digit is displayed by seven segments.
Range switching Automatic and manual
Input method Floating
Measurement method Integration
Overload display OL
Trigger function External trigger (7351A, 7351E + Option 03)
Memory Data memory: Up to 20,000 data
Condition setting memory: 4 (USER0 to USER3)
Power supply AC power supply: 100 V/120 V/220 V/240 V
(Can be switched by user)
Specify the option when ordering.
Use a power cable and fuse that are compliant with the safety standard when
changing the power supply voltage.
Power supply frequency 50 Hz/60 Hz
Power consumption 20 VA or below
Dimensions Approx. 212 (width) 88 (height) 340 (depth) mm
Mass 3.4 kg or less
Safety Compliance with IEC61010-1 Ed.2, Category II
EMI Compliance with EN61326 classA
Compliance with FCC classA
Option number Standard OPT.32 OPT.42 OPT.44
Power supply voltage 100 V 120 V 220 V 240 V