EasyManua.ls Logo

ADCMT 7351A - Page 165

ADCMT 7351A
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7351 Series Digital Multimeter Operation Manual
8.3.4 Calibration Procedure
8-9
7. Pressing the key displays the measurement value as a setting value.
8. Adjust the setting value to the calibration value by using the , , , and
keys.
NOTE: Press the key if calibration must be cancelled due to a mistake
such as entering an incorrect number.
9. Press the key to execute the calibration.
10. Repeat steps 6 to 9 when calibrating other ranges.
Calibrating the full-scale point
11. Connect the calibration standard to the input terminal.
12. Set a range to be calibrated.
For multiple calibration points, calibrate them in the order of calibration point 1,
2, and 3.
13. Input a voltage equivalent to the calibration point.
14. Pressing the key displays the measurement value as a setting value.
15. Adjust the setting value to the calibration value by using the , , , and
keys.
NOTE: Press the key if the calibration must be cancelled due to a mistake
such as entering an incorrect number.
16. Press the key to execute the calibration.
17. Repeat steps 12 to 16 when calibrating other ranges.
18. Repeat steps 4 to 17 when calibrating other functions.
Canceling the calibration mode
19. Select OFF from ‘CAL’ in in .
20. Press to determine the setting.
The CAL indicator is turned off and indicates that the calibration mode is can-
celed.
21. Press to exit from MENU.
HOLD
EXIT
ENTER
HOLD
EXIT
ENTER
9 SYS
MENU
ENTER
EXIT

Table of Contents

Related product manuals