156 1260 Infinity Binary Pump User Manual
10 Maintenance
Exchanging the Purge Valve Frit or the Purge Valve
5 Place a new frit into the purge valve with the orientation of the frit as
shown below (slit in frit points to the front). Reinstall the seal cap
including the gold seal.
6 Insert the purge valve into the purge valve holder and orient the waste
outlet nozzle downward.
7 Tighten the purge valve and reconnect outlet capillary and waste
tubing.
NOTE
Before reinstallation always check the gold seal in the seal cap. A deformed seal cap
should be exchanged.