To customize your analyzer
Creating an AUTO_ST.STA file allows you to automatically specify a specific
measurement state when you turn on the analyzer or when you press [
Preset
]
[
RECALL AUTOSTATE
].
1 Initialize the analyzer.
Press [
Preset
] [
DO PRESET
].
2 Configure the analyzer to fit your application.
3 Select the default disk, either the non-volatile RAM disk or the internal disk.
See To select the default disk in the second section, Setting Up.
4 Save the measurement state.
If the internal disk drive is the default disk, insert a 3.5 inch floppy disk.
Press [
Save/Recall
] [
SAVE MORE
][
SAVE AUTOSTATE
].
Press [
ENTER
].
At power-up, the analyzer searches the internal disk drive first and then searches the
non-volatile RAM disk for the AUTO_ST file. If it is found, the analyzer loads the
measurement state immediately.
Agilent 35670A Quick Start Guide Automating Measurements