10 Maintaining the VI
To Install a Capillary Column with the Volatiles Interface
170 Maintaining Your GC
20 After the GC becomes ready, wait 10 minutes then ignite the detector flame or adjust
offset on the NPD bead.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
21 Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
the fittings.
22 Reset the EMF counter.
23 Perform an inlet Leak & Restriction test and reset the maintenance counters.