EasyManua.ls Logo

Agilent Technologies 8890 - Page 80

Agilent Technologies 8890
282 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Maintaining the Purged Packed Inlet
To Install a Capillary Column with the Purged Packed Inlet
80 Maintaining Your GC
11 Adjust the column position so that the septum is even with the bottom of the column nut.
Finger-tighten the column nut until it begins to grip the column.
12 Tighten the column nut an additional 1/4 to 1/2 turn with a wrench so that the column
cannot be pulled from the fitting with gentle pressure.
13 Plug in the new column’s Smart ID key. See Figure 25.
Figure 25. Insert Smart ID key
14 Configure the new column.
15 Condition the column per the manufacturer’s recommendation. See “To Condition a
Capillary Column”.
16 Install the column into the detector.
To Install a Capillary Column in the FID
To Install a Capillary Column in the NPD
To Install a Capillary Column in the TCD
To Install a Capillary Column in the ECD
To Install a Capillary Column to the FPD+
17 Exit maintenance mode. Select Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
Smart ID key

Table of Contents

Other manuals for Agilent Technologies 8890

Related product manuals