EasyManua.ls Logo

Agilent Technologies 990 - Page 131

Agilent Technologies 990
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9Errors
Error List
Agilent 990 Micro GC User Manual 131
55 INFO One cycle of flush passed
56 WARN Flush cycles aborted
57 INFO Start column clean
58 INFO Column clean passed
59 WARN Column clean aborted
60 INFO Temperature equilibrating after column cleaning
61 INFO Temperature equilibrating passed
62 INFO TCD calib, chan = <n>
63 INFO TCD calib success, chan = <n>
64 RECOVERABLE TCD calib failed, chan = <n>
65 WARN Detector ADC error, chan = <n>
66 INFO Detector enabled, chan = <n>
67 INFO Detector disabled, chan = <n>
68 WARN TCD temperature limit activated, chan = <n>
69 WARN TCD temperature limit deactivated, chan = <n>
70 WARN Channel unit changed, chan = <n>
71 CRITICAL ADC mux offset Is out of range, chan = <n>
72 INFO ADC mux offset is ok, chan = <n>
73 WARN Stream select failed
74 INFO Stream select ok
75 RECOVERABLE Ambient temperature too high
76 INFO Ambient temperature error cleared
77 RECOVERABLE Ambient pressure too high
78 INFO Ambient pressure error cleared
79 WARN Battery 1 lower power
80 INFO Battery 1 power restored
81 WARN Battery 2 lower power
82 INFO Battery 2 power restored
83 RECOVERABLE Low power supply
84 INFO Main power supply restored
85 CRITICAL Internal power failure, chan = <n>
86 INFO Internal power failure restored. chan = s<n>
87 INFO Loading mainboard EDS
88 INFO Loading channel controller <n> EDS
89 INFO Loading analytical module <n> EDS
90 WARN Mainboard EDS logging error
91 WARN CCB EDS logging error, chan = <n>
Table 31 Error list (continued)