4Sample Gas Handling
G3588-68864
Agilent 990 Micro GC User Manual 55
6 Connect the 1/16 tubing to the sample inlet manifold by the 1/16 inch Swagelok fitting.
WARNING
The pressure regulator has a maximum inlet pressure of 3,000 psi. Applying higher pressures
may result in serious personal injury and equipment damage.
7 Connect the Sample IN port on the pressure regulator to the sample input line.
8 Start the GC (see Long Storage Recovery Procedure on page 41).
9 Leak test the system to ensure that all connections are leak free.