EasyManuals Logo

Agilent Technologies 990 User Manual

Agilent Technologies 990
170 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
5 Micro GC Channels
Backflush Option
76 Agilent 990 Micro GC User Manual
When all compounds to be quantified are transferred to the analytical column, the backflush valve
switches (at the backflush time). On the precolumn, the flow inverts and all compounds left on the
precolumn now backflush to the vent. On the analytical column the separation continues because
there the flow is not inverted. See Figure 28.
The standby mode is the backflush configuration (if the instrument is equipped with the optional
backflush valve).
Backflushing saves the time required to elute high boiling components that are not of interest, and
ensures that the precolumn will be in good condition for the next run.
Figure 28. Backflush flows
Precolumn
Analytical
column
Detector
Injector
Backflush
vent
Restriction
Pressure
regulator
System
pressure
Pressure point

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 990 and is the answer not in the manual?

Agilent Technologies 990 Specifications

General IconGeneral
BrandAgilent Technologies
Model990
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals