128 Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
7 LAN Communication, Installation, Connection and Configuration
Using the Agilent BootP Service
The entries are described below
#8454 lab
Comment line starting the configuration set for an Agilent Cary 8454
UV-Visible spectrophotometer. A comment is preceded with #; all
characters after the # on the same line are ignored by the BootP
Service.
UV-Vis:\
Host name of the instrument. The host name communication can only
be used if a DNS server (Domain name server) is correctly set up for
the name or a corresponding entry in the HOSTS file exists (windows\
system32\drivers\etc\hosts).
ha=0030D30D3277:\
This is the hardware or MAC address of the G1369C LAN card. It is
usually printed on a label attached to the card. If not, see section
“Determine your MAC Address” on page 123.
ip=<IP address - ask your system administrator>
This is the IP address that will be assigned to the card after the BootP
request. This IP address must be configured using the Agilent
ChemStation Configuration Editor.
sm=255.255.255.0:\
This is the subnetmask assigned by the BootP request.
gw=<gateway address - ask your system administrator>
This is the gateway address assigned by the BootP request.
T145=64:T146=01:T147=01
These parameters configure the buffer handling of the G1369C LAN
card. These are required for the instrument on the LAN to work
properly and must not be changed.
Different entries for the subnet mask and default gateway might be
required, depending on your network setup. These values need to be set
by your network administrator.

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals