Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide 127
LAN Communication, Installation, Connection and Configuration 7
Using the Agilent BootP Service
Configuring the G1369C LAN Card Using BootP
In order for an internal G1369C LAN card to use BootP, the default
setting: bootp=YES must be set. For more details please see the user
manual of your G1369C LAN interface card.
To change the setting, the card must be reset; this is done each time the
instrument with the G1369C LAN card inserted is power cycled.
The Agilent BootP Service uses a file called bootptab were all the
configuration information is stored. The BootP Service program on the
Agilent ChemStation CD- ROM allows easy editing of the bootptab file. If a
BootP Service is already installed, the following information must be
supplied to the system administrator to be able to correctly set up the
bootptab file.
# CAG Bootptab file
# global defaults
global.defaults:\
sm=255.255.255.0:\
gw=<gateway address - ask your system administrator>
#8454 lab
UV-Vis:\
ht=1:\
ha=0030D30D3277:\
ip=<ip address - ask your system administrator>
sm=255.255.255.0:\
gw=<gateway address - ask your system administrator>
vm=auto:\
hn:\
bs=auto:\
T145=64:T146=01:T147=01

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals