EasyManuals Logo

Agilent Technologies Cary 8454 Installation Guide

Agilent Technologies Cary 8454
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #56 background imageLoading...
Page #56 background image
56 Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
4 Installing Accessories
Cell Stirring Module
Tubing Connections and Operation
The cell stirring module can be driven by either water or air. Water is
preferable because it is more easily controlled. Water temperature can be
up to 100 °C (212 °F) with a maximum allowable back pressure of 4 bar
(approximately 60 psi). If air is used to drive the cell stirring module, the
pressure should be less than 0.3 bar (approximately 5 psi) when the
stirrer is started. Again, the maximum pressure is 4 bar (approximately
60 psi).
There are several ways you can connect a water or air supply to the cell
stirring module, see Figure 16. In all cases we recommend using 3/16 inch
i.d. silicon tubing between the water supply and the stirring module.
The stirring module can be used alone, without the thermostattable
functions of the cell holder.
The same water supply can be used to regulate sample cell temperature
and to power the cell stirring module.

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals