Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide 111
GPIB Communication, Connections and Configuration 6
Changing the GPIB Address of the Spectrophotometer
GPIB Address Settings
Use Table 5 and Table 6 to find the switch settings for the GPIB address
of your spectrophotometer.
Table 5 GPIB address switch settings (addresses 0 through 15)
Address 1234 5678
0 00000000
1 00000001
2 00000010
3 00000011
4 00000100
5 00000101
6 00000110
7 00000111
8 00001000
9 00001001
10 00001010
11 00001011
12 00001100
13 00001101
14 00001110
15 00001111
Switches 1 and 2 are used for factory firmware settings. Do not change these
switches. Any changes could result in instrument failure. To recover, set switches 1
and 2 to position 0 and power-cycle the instrument. For more detailed information,
see the Service Manual.