110 Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
6 GPIB Communication, Connections and Configuration
Changing the GPIB Address of the Spectrophotometer
3 Remove any objects (for example, cuvettes, flow cells) from the light
path and turn on the line power to the spectrophotometer. This action
stores the GPIB address in the non- volatile memory.
The address will remain unchanged in the non- volatile memory until a
new address is set on the DIP switches and the instrument is
power- cycled again.
CAUTION
Make sure that the GPIB address does not conflict with other instruments connected
to the same GPIB controller and the GPIB address of the controller itself.
In our configuration the GPIB address 30 is used for the controller.