EasyManuals Logo

Agilent Technologies Cary 8454 Installation Guide

Agilent Technologies Cary 8454
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #20 background imageLoading...
Page #20 background image
20 Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
2 Installation and Start Up
Site Requirements for Your Spectrophotometer
Power cords
Different power cords are offered as options with the spectrophotometer.
The female end of each of the power cords is identical. This connects to
the power-input socket at the rear of the spectrophotometer. The male end
of each of the power cords is different and designed to match the wall
socket of a particular country or region.
Do not position the equipment so that it is difficult to operate the
disconnecting device.
Space
The spectrophotometer is 34.4 cm (13.5 inches) wide, 56.0 cm
(22.0 inches) deep and 18.5 cm (7.3 inches) high (see Figure 1 on
page 24). It weighs 14 kg (30.86 lbs) and will fit on almost any desk or
laboratory bench. It needs an additional 2.5 cm (1.0 inches) of space on
either side to allow for the circulation of air for cooling, and
approximately 8 cm (3.1 inches) in the rear for electrical connections.
The computer, including the keyboard, is approximately 50 cm
(19.7 inches) wide and 60 cm (23.6 inches) deep.
Additional space is required for accessories like pumps and autosamplers.
NOTE
Do not replace the mains power cord with one that is inadequately rated. When replacing
the mains power cord, use only a mains power cord with a rating equivalent to the one
supplied with the instrument. See the instrument installation guide for detailed mains
power cord rating information.
WARNING
Always operate your instrument from a power outlet which has a ground
connection. Always use the power cord designed for your region.

Table of Contents

Other manuals for Agilent Technologies Cary 8454

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Cary 8454 and is the answer not in the manual?

Agilent Technologies Cary 8454 Specifications

General IconGeneral
BrandAgilent Technologies
ModelCary 8454
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals