Triggers 10
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 157
Nth Edge Burst trigger set up consists of selecting the source, the slope of
the edge, the idle time, and the number of the edge:
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
to select Nth Edge Burst.
3 Press the Source softkey, and turn the Entry knob to select the input
channel source.
4 Press the Slope softkey to specify the slope of the edge.
5 Press the Idle softkey; then, turn the Entry knob to specify the idle
time.
6 Press the Edge softkey; then, turn the Entry knob to which edge number
to trigger on.