EasyManua.ls Logo

Agilent Technologies InfiniiVision 3000 - Page 174

Agilent Technologies InfiniiVision 3000
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
174 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
10 Triggers
5 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
to select USB.
6 Press the Trigger: softkey to select where the USB trigger will occur:
SOP (Start of Packet) — triggers at the Sync bit at the beginning of
the packet.
EOP (End of Packet) — triggers at the end of the SE0 portion of the
EOP.
RC (Reset Complete) — triggers when SE0 is > 10 ms.
Suspend (Enter Suspend) — triggers when the bus is idle > 3 ms.
Exit Sus (Exit Suspend) — triggers when exiting an idle state > 10 ms.
This is used to see the suspend/resume transition.
7 Press the Speed softkey to select the speed of the transaction being
probed.
You may select Low Speed (1.5 Mb/s) or Full Speed (12 Mb/s).
8 Press the D+ and D– softkeys to select the channel connected to the
USB signal D+ and D– lines. The D+ and D– labels for the source
channels are automatically set.

Table of Contents

Related product manuals