238 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
14 Measurements
 
To reset the statistics measurements, press the Reset Statistics softkey. This 
resets all statistics and begins recording statistical data again.
Each time a new measurement (for example: frequency, period, or 
amplitude) is added the statistics are reset and accumulation of statistical 
data begins again.
When the [Single] key is pressed, statistics are reset and a single 
measurement is done (count = 1). Successive [Single] acquisitions 
accumulate statistical data (and the count is incremented).
Press the Transparen t softkey to disable the Transparent mode. This shows 
statistics with a gray background. Press the Transparent softkey again to 
enable Transparent mode. This writes measurement values, statistics, and 
cursor values on the screen without a background. The Transparent 
setting affects measurement statistics, reference waveform information, 
and the optional Mask test feature's statistics display.
Relative σ — When enabled, the standard deviation shown in measurement 
statistics becomes a relative standard deviation, that is, standard 
deviation/mean.
Max Count — This softkey specifies the number of values used when 
calculating measurement statistics.
The Increment Statistics softkey only appears when the acquisition is stopped 
and the optional segmented memory feature is off. Press the [Single] or 
[Run/Stop] key to stop the acquisition. You can use the horizontal position 
control (in the Horizontal control section of the front panel) to pan 
through the waveform. Active measurements will stay on screen, allowing 
you to measure various aspects of the captured waveforms. Press Increment 
Statistics to add the currently measured waveform to the collected 
statistical data.
The  Analyze Segments softkey only appears when the acquisition is stopped 
and the optional segmented memory feature is on. After an acquisition 
has completed (and the oscilloscope is stopped), you can press the Analyze 
Segments softkey to accumulate measurement statistics for the acquired 
segments.
You can also turn on infinite persistence (in the Display Menu) and press 
the  Analyze Segments softkey to create an infinite persistence display.