EasyManuals Logo

Agilent Technologies InfiniiVision 3000 User Manual

Agilent Technologies InfiniiVision 3000
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #328 background imageLoading...
Page #328 background image
328 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
21 Reference
Waveform Data
Header
A waveform may have more than one data set. Each waveform data set
will have a waveform data header. The waveform data header consists of
information about the waveform data set. This header is stored
immediately before the data set.
X Display Origin A 64-bit double that is the X-axis value at the left edge of the display. For time
domain waveforms, it is the time at the start of the display. This value is
treated as a double precision 64-bit floating point number. If the value is zero
then no data has been acquired.
X Increment A 64-bit double that is the duration between data points on the X axis. For
time domain waveforms, this is the time between points. If the value is zero
then no data has been acquired.
X Origin A 64-bit double that is the X-axis value of the first data point in the data
record. For time domain waveforms, it is the time of the first point. This value
is treated as a double precision 64-bit floating point number. If the value is
zero then no data has been acquired.
X Units A 32-bit integer that identifies the unit of measure for X values in the acquired
data:
0 = Unknown.
1 = Volts.
2 = Seconds.
3 = Constant.
4 = Amps.
5 = dB.
6 = Hz.
Y Units A 32-bit integer that identifies the unit of measure for Y values in the acquired
data. The possible values are listed above under X Units.
Date A 16-byte character array, left blank in InfiniiVision oscilloscopes.
Time A 16-byte character array, left blank in the InfiniiVision oscilloscopes.
Frame A 24 byte character array that is the model number and serial number of the
oscilloscope in the format of: MODEL#:SERIAL#.
Waveform Label A 16 byte character array that contains the label assigned to the waveform.
Time Tags A 64-bit double, only used when saving multiple segments (requires
segmented memory option). This is the time (in seconds) since the first
trigger.
Segment Index A 32-bit unsigned integer. This is the segment number. Only used when saving
multiple segments.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals