FlexRay Triggering and Serial Decode 23
Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide 359
2 In the Search Menu, press the Search softkey; then, turn the Entry knob
to select the serial bus (Serial 1 or Serial 2) on which the FlexRay
signals are being decoded.
3 In the Search Menu, press Search for; then, select from these options:
Frame ID — Finds frames with the specified ID. Press the Frame ID
softkey to select the ID.
Cycle number (+ Frame ID) — Finds frames with the specified cycle
number and ID. Press the Frame ID softkey to select the ID. Press
the Cycle # softkey to select the number.
Data (+ Frame ID + Cycle number) — Finds frames with the specified data,
cycle number, and frame ID. Press the Frame ID softkey to select the
ID. Press the Cycle # softkey to select the number. Press the Data
sofkey to open a menu where you can enter the data value.
Header CRC Error — Finds cyclic redundancy check errors in headers.
Frame CRC ErrorFinds cyclic redundancy check errors in frames.
Errors — finds all errors.
For more information on searching data, see "Searching Lister Data" on
page 126.
For more information on using the [Navigate] key and controls, see
"Navigating the Time Base" on page 58.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfiniiVision 3000 and is the answer not in the manual?

Agilent Technologies InfiniiVision 3000 Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfiniiVision 3000
CategoryTest Equipment
LanguageEnglish

Related product manuals