382 Agilent InfiniiVision 3000 X-Series Oscilloscopes User's Guide
25 I2S Triggering and Serial Decode
 
6 For the SCLK (serial clock), WS (word select), and SDATA (serial data) 
signals:
a Connect an oscilloscope channel to the signal in the device under 
test.
b Press the SCLK,  WS, or SDATA softkey; then, turn the Entry knob to 
select the channel for the signal.
c Press the corresponding Threshold softkey; then, turn the Entry knob 
to select the signal threshold voltage level.
Set the threshold levels for the SCLK, WS, and SDATA signals to the 
middle of the signals.
The threshold voltage level is used in decoding, and it will become 
the trigger level when the trigger type is set to the selected serial 
decode slot.
The SCLK, WS, and SDATA labels for the source channels are 
automatically set.
7 Press the  Back/Up key to return to the Serial Decode Menu.
8 Press the Bus Config softkey to open the I
2
S Bus Configuration Menu 
and display a diagram showing WS, SCLK, and SDATA signals for the 
currently specified bus configuration.
9 Press the Word Size softkey. Turn the Entry knob to match the 
transmitter word size of the device under test (from 4 to 32 bits).
10 Press the Receiver softkey. Turn the Entry knob to match the receiver 
word size of the device under test (from 4 to 32 bits).
11 Press the Alignment softkey; then, turn the Entry knob to select the 
desired alignment of the data signal (SDATA). The on- screen diagram 
changes with your selection.