124 Agilent 1290 Infinity II Flexible Pump User Manual
7 Maintenance
Pump Head Procedures
3
Assemble the pump head by putting the pump chamber
housing on top of the support ring. Observe correct
orientation of the pin.
NOTE
Do NOT install the inlet and outlet valves and the
outlet filter at this stage.
4
Loosely close the 4 screws at the rear of the pump head.
The screws will be fixed tightly later.
E^c
EjbeX]VbWZg]djh^c\
Ejbe]ZVY[ZggjaZh
Hjeedgig^c\
E^hidc]djh^c\