Installation 3
Agilent SPS 4 Autosampler User’s Guide 43
6 Slip the tip of the sample probe into the hole in the middle of the
Z- axis slide, through the mount nut and the accessories plate until the
probe mount block is sitting in the z- axis slide.
7 Slide the mount nut up onto the probe mount block and tighten to
secure the sample probe to the Z- axis slide.
8 Route the sample line from the sample probe to the left around the
probe arm housing. When using a carbon fiber sample probe, do not
run the sample line through the hook (or tube restraint) at the top of
the Z- axis slide. Connect the cable tie to the eyelet on the underside of
the probe arm cover and leave a loop of approximately 10 mm. Thread
the sample line through the loop in the cable tie as shown in Figure 14.
Figure 14 Routing the sample line tubing
NOTE
Do not use the tube restraint for pre- formed sample probes. It is only
used for non pre- formed sample probes.
Cable tie tubing through
the eyelet on bottom of
housing
Tube Anchor Grommet
Colored Markers