VS Series Helium Mass Spectrometer Leak Detectors
DRAFT 4/23/15
2-31
Ion Source Filament
See Section 3.1.2 “Manual Tuning”
on page 3-7.
Filament 1 Off Filament 1 is selected and not lit.
Filament 1 On Filament 1 is selected and lit.
Filament 2 Off Filament 2 is selected and not lit.
Filament 2 On Filament 2 is selected and lit.
Gain Gain (Gain Value) Displays the gain that applied to the
leak rate. The actual helium signal is
multiplied by the gain value to read
the correct, calibrated leak rate.
Leak Rate Range
See Section 2.2.4.4 “Leak Rate
Ranging” on page 2-25.
Auto Ranging System is set to auto range through
the
leak rate scale.
Manual Ranging System is set for manual range control.
Range Stop
See Section 2.2.4.4 “Leak Rate
Ranging” on page 2-25.
Range Stop -[Exp] System is set to display the most
sensitive leak rate range, 10
-
[Exp]
.
The
range stop line shows only if
range stop is on.
Reject Set Point
See Section 2.2.4.2 “Reject and
Audio Set Points” on page 2-19.
Reject# 1,2,3,4 Active Reject set points displayed are active;
this line is blank when all set points
are disabled or inactive.
Software Version and
Agilent Web site
Shows the software version of the CPU
board (L version) and software version
of the display board (D
version). Also
shows the Agilent web address for
customer use.
See the back cover of this manual for a
listing of our sales and service offices.
Split Flow
See Section 2.2.4.1 “Calibration
Set-Up” on page 2-17.
Shows split flow, only if the feature is
turned on. To change the status, go
to
the Advanced Parameters screen.
Vent Lockout Status
See Section 2.2.4.7 “Control Panel
Access” on page 2-32.
Shows the status of the vent lockout
feature only if it is on. To c hange the
status of the vent lockout feature, go
to
the Control Panel access screen.
Table 2-6 System Information Screen Conditions (Continued)
Condition Display Description

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies VS Series and is the answer not in the manual?

Agilent Technologies VS Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelVS Series
CategorySecurity Sensors
LanguageEnglish

Related product manuals