Chapter 3 Unpack and Installation
13
Chapter 3 Unpack and Installation
Guide:
Installation Environment
3.1 Installation Environment
To select installation position, take the following into consideration:
1 Away from flammable, explosive or corrosive substance, e.g., alcohol, thinner, sulfuric acid
etc.
Operating Temp.: 0℃~+40℃
Storage Temp.: -10℃~+50℃
Avoid sudden change of temperature, otherwise condensate will occur inside the instrument.
3 Away from boilers, humidifiers and water and so on.
Operating relative humidity: 20~75%RH
Storage relative humidity: 0~90%RH
Stop usage of the tester as condensate occurs.
4 Away from strong electromagnetic interference.
5 Away from the obvious shock and vibration.
6 Keep the working environment free of dust. Keep good ventilation. The scanner is cooled via
natural air. Poor ventilation can lead to damage to the equipment. Keep the rear panel at least
30cm away from the wall during operating.
7 Away from precise instruments - As high-voltage is output, corona discharging will occur near
the test point of DUT, emitting radio electromagnetic wave and interfering measuring
accuracy of precise instruments.
3.2 Unpack and Check
First check the nameplate and see if the model is consistent with that in the order; then check
if the items are complete according to the Packing List in the manual. If the items are not
consistent with those in the Packing List, please contact Ainuo service center or the dealer.
If the package is damaged, check the instrument for any deformation or scratch, or damage to
the panel etc. In case of any damage, inform Ainuo Instrument Company or the dealer
immediately. Our service center will repair or replace it with new one. Do not return the product
before informing Ainuo Instrument Company or its dealer.