EasyManua.ls Logo

Ainuo AN9640B - System Settings; Alarm Volume; Password

Ainuo AN9640B
81 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Settings and Test
27
No.
F menu
Function & Description
1
Start Test
Enter into test standby page
2
Group Selection
Enter into Group selection module to select and call the group
3
Parameters Settings
Enter into Parameters setting module to set the current selected group
4
System Settings
Enter into System setting module to set system parameters
5
Extended Function
Enter into Extended Function module to set barcode,U-disk, channel scan,
print, PLC etc information
6
Calibration
Enter into Calibration module to make calibration or reset factory settings
etc.
For first power-on, it is recommended to set the analyzer as the following sequence.
4.3 System Settings
System setting screen is as shown in Figure 4-3-1.
Figure 4-3-1 Analyzer Settings
Press or to move the cursor, press F menu button and numeric key to set
parameters in system setting page.
After finishing settings, press Exit to return main page, select “Save” or “Cancel” current
settings and return to the main page.
Caution! Be careful to change the settings of the instrument.
The settings of various parameters are described in the following sections.
4.3.1 Alarm Volume
To set the buzzer alarm sound, including 1 to 9 levels. Level-0: closed.
4.3.2 Password
1) Password function: ON/OFF. If the password is set to ON, before enter into Parameters
Settings page, System Settings and Extended Function in the Function Selection page, the

Table of Contents

Related product manuals