Chapter 1 Safety Rules
2
1.3 Testing Station
1 Position
The station must be positioned away from persons other than the operators. Keep the station
isolated with other facilities, indicating “High-voltage Testing Station”. Indicate
“Danger! Testing ..., Non-operators keep away!” during testing.
2 Power supply
Available power supply: single-phase 220V±10%, 50Hz±5%. Ensure proper power supply
to avoid damage to instrument and personal injury.
Only use fuse of proper specification.
Ground the tester well before operation so as to guarantee personnel safety. Set a power
switch for the testing station at the entrance and prompt especially. In case of emergency, shut off
the power supply and then handle the accident.
3 Workplace
Use non-conductive worktable or platform as possible. Metal between the operator and DUT
is forbidden. The operators shall not operate or adjust the tester over the DUT. The testing place
must be clear, clean and in order. Keep the instrument and the testing line in proper place after
operation. All persons shall identify the objects under testing, to be tested and that tested. Keep the
testing station and surrounds free of flammable or corrosive gases. Do not operate the tester near
flammable substance.
1.4 Operating Regulation
1 Qualification
Any electric shock due to wrong operation of the voltage and current from this analyzer will
cause injury to persons, or even death. This analyzer must be used and operated by qualified
operators according to the manual.
2 Safety rules
The operators must be educated and trained in time, so as to understand all operating
regulations, and operate this analyzer taking the safety rules into consideration.
3 Clothing
The operators shall not wear clothing with metal garnishry or metal jewelry and watches, to
avoid electric shock or even more serious results.
The operators must wear insulated gloves during operation of the analyzer.
4 Medical regulation
This analyzer can not be operated by persons wearing cardiotach adjuster or pacemaker.