Chapter 1 Safety Rules
3
1.5 Operating Rules
The maximum output of this tester is 5kVAC. Observe the
following provisions during test to avoid injury or death!
1.5.1 Forbidden
Frequent ON/OFF of the instrument is forbidden. Switch on/off the instrument at interval
of 30 seconds or move.
Never open the house of the analyzer. The analyzer must be maintained by trained and
qualified engineers or technicians.
1.5.2 Precautions during testing
The operators must wear insulated gloves.
To leave the operating area temporarily or to perform the test later, switch off the power
supply.
During high-voltage testing, never touch DUT or any objects connected with the DUT.
The operator must be able to master the control switch and the remote switch
independently. Please remove the remote switch if it is not used. Non-qualified operators
and other persons should keep away from high voltage area.
Connect the earth wire of analyzer properly. Only insert the high-voltage testing line for
testing. Hold the insulated parts of high-voltage line (clip). Never directly touch the
high-voltage output terminal (clip).
In case of any problems, press STOP immediately. Turn off the power supply directly if necessary.