Annex C Protocol
71
Return DCW, 1.50kV, 0.0uA, OK ,;
1) FN name: create a file.
“name” is the file name, up to 30 characters. This file is saved directly in the file location of
the current call.
2) Add test items (total 8): SET-ACW, SET-DCW, SET-IR, SET-GB, SET-TCT, SET-PW,
SET-ST, SET-WAIT.
Arguments:
1、 Arguments are separated by commas, arranged in chronological order in the following
table.
2、 The number of arguments is free (0 to the Max.).
For example: If no argument is set, all arguments are default;
In case 4 arguments are set, the first four arguments are the settings, and the follow-up
arguments are default.
In case the number of arguments exceeds the maximum, the front ones prevail, and the
subsequent arguments are invalid.
The arguments must be consecutive. To keep any middle argument the default one, set it
to -1 (-1 = default).
3、 For any argument, if the setting is outside the range, giving error ExceedPara, and this
command is invalid.
The units of arguments are fixed, and the command does not contain units. Refer to the
following table.
First 4 key arguments
Name of arguments, setting range, default
1) Output Voltage,100~5000V,1500V
2) Current Upper Limit,0.00~100.00 mA,3.5mA
3) Current Lower Limi,0.000~9.999mA,0mA
4) Test Time,0.5~999.9s(0=LLLL),1.0s
5) Three-channel scanning option, 0 \ 1 \ 2,0 represents input to output, 1
represents input to ground, 2 represents output to ground
6) Ramp Time,0.1~999.9s(0=OFF),0.1s
7) Descent Time,0.1~999.9s(0=OFF),0s
9) Compensate Switch,0~1,0 (0=OFF,1=ON)
10) Output Frequency,0~1,0 (0=50Hz,1=60Hz)