Alarm Condition / Description / Corrective Action
Flow Out of Tolerance*
Sample ow rate too high or too low.
Corrective Action:
Check the Inlet and Outlet pressure. Call AMETEK Service if the problem persists.
Battery Low*
Battery needs to be replaced.
Corrective Action:
Call AMETEK Service.
Reference Gas*
Analyzer detected problem with Reference gas.
Corrective Action:
Check and/or replace the Dryer. Call AMETEK Service if the problem persists.
Enclosure Temperature
Excessive internal temperature.
Corrective Action:
External temperature should be 50 °C or less. Call AMETEK Service.
Moisture Generator, Date N/A
Moisture generator date has expired.
Corrective Action:
Replace the Moisture Generator.
Dryer Alarm
Dryer failure Imminent.
Corrective Action:
Replace the Dryer.
Reading Alarm
Moisture concentration is out of user dened limits, or analyzer is Oine verifying, or pressure
is outside of range for dewpoint calculations.
Corrective Action:
Review alarm settings and use the Congurator Software to identify source of error.
Data Valid contact opens on all alarms and stays closed during
normal functions and readings. An open Data Valid contact indicates
Verication is in process or an alarm condition.
Maintenance and Troubleshooting | 4-23