Mechanical Installation
Install the 3050-OLV Analyzer as close as possible to the sample source. The
unit should be protected from direct exposure to weather and located so that
the ambient temperature specications will not be exceeded (see Figure 2-2
for connection locations).
1. If not already installed, install a main process shut-o valve and pressure
reducer (recommended) at the Sample tap (Figure 2-3).
2. Connect the Dryer input to the To Dryer” 1/8-inch compression tting on
the analyzer. Connect the Dryer output to the “From Dryer 1/8-inch com-
pression tting. Refer to “Dryer Installation Instructions” and Figure 2-4.
3. Connect the exhaust tting to appropriate vent system.
Leave “Exhaust Out” tting capped or blocked with an isolation valve
until Sample gas is owing. This prevents ingress of wet ambient air.
“Sample In” is protected with internal valve, which is closed when
power is O.
4. Open the main process shut-o valve and purge sample line to an ap-
propriate area for at least ve (5) minutes. Close the main process shut-o
valve. This will help prevent contamination from entering the sensor.
5. Connect the sample line to the sample inlet 1/8-inch compression tting.
2-6 | 3050-OLV Moisture Analyzer

Table of Contents

Other manuals for Ametek 3050-OLV

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek 3050-OLV and is the answer not in the manual?

Ametek 3050-OLV Specifications

General IconGeneral
BrandAmetek
Model3050-OLV
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals