Important Notice to Users
SPECIAL WARNINGS AND INFORMATION FOR USE OF THIS EQUIPMENT IN CLASS I,
DIVISION 2 HAZARDOUS LOCATIONS:
This Equipment is designed to meet the requirements for: Class I, Division 2, Groups
ABCD, T4 or Non-Hazardous Areas Only. (Note: The Division 2 Standalone Analyzer is
not supplied with the 24 VDC power supply option.)
Explosion Hazard – Substitution of Components May Impair Suitability for
Class I, Division 2.
AVERTISSEMENT – RISQUE D’EXPLOSION – LA SUBSTITUTION DE COM-
POSANTS PEUT RENDRE CE MATERIEL INACCEPTABLE POUR LES EMPLACE-
MENTS DE CLASSE I, DIVISION 2.
Explosion Hazard – Do not disconnect equipment unless power has been
switched o or the area is known to be non-hazardous.
AVERTISSEMENT – RISQUE D’EXPLOSION – AVANT DE DECONNECTER
L’EQUIPEMENT, COUPER LE COURANT OU S’ASSURER QUE L’EMPLACEMENT EST
DESIGNE NON DANGEREUX.
All Input and Output Wiring Must be in Accordance with Class I, Division 2
Wiring Methods (NEC Sec 501.4(b) or CEC 18-152) and in Accordance with the
Authority Having Jurisdiction.
If the 3050-OLV is to be powered by a source of 24 VDC other than that supplied by
AMETEK, the power sources output must be isolated from hazardous mains voltages
by use of double or reinforced insulation, having a minimum dielectric strength
2300VAC. When the 3050-OLV is used in a Class I, Division 2 area, this external power
source must be located in a General Purpose area or be Division 2 approved.