PN 305200901, Rev YT
Specication Description
Voltage and Power
Requirements
Base Analyzer: 24 VDC, 50 watts maximum
Analyzer with
Enclosed Sample System: 120 ±10 % VAC, 50/60 Hz, 150 W
maximum
230 ±10 % VAC, 50/60 Hz, 150 W
maximum
Moisture Generator Value 50 PPMV nominal
Weight of Analyzer 4.2 Kg
Minimum PC
Requirements
Windows XP, 7, or 10
3050-OLV Moisture Analyzer Approvals and Certications
Approvals and
Certications
UL/CSA General Safety Requirements (General Purpose)
UL/CSA Class I, Division 2, Groups A, B, C, D T4
UL/CSA Class I, Division 1, Groups B, C, D T5 T6
ATEX/IECEx db eb IIB + H2 T* Gb IP65
T6: -20 to 40 °C
or
T5: -20 to 50 °C
Complies with all relevant European directives:
EAC Ex db eb IIB + H2 Gb
GOST R Pattern Approval
Environmental Pollution Degree 2 (all Analyzer versions)
Installation Category II (all Analyzer versions)
Indoor use only (base Analyzer)
3050-OLV Moisture Analyzer Rating Label
Process Control
Equipment for
Hazardous
Locations
87GL
LVD
EMC
Class A, Group 1
305453001
WARNING
5-2 | 3050-OLV Moisture Analyzer