PN 305200901, Rev YT
Working From the Verication Tab
Use the Verication tab to schedule a routine Verication.
Figure 3-6.1.
Verication tab.
Verify Now
Click this button to start the Verication.
Abort
Click this button to terminate the Verication.
Calibrate after verify
Checked The analyzer performs a Span adjustment at
the end of the Verication.
Not Checked The analyzer performs only a Verication.
Ignore span drift
Checked One time Span value change is not limited.
Not Checked One time Span value change is limited to 10 %
of the existing Span value.
3-14 | 3050-OLV Moisture Analyzer