EasyManua.ls Logo

Ametek 3050-OLV

Ametek 3050-OLV
103 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PN 305200901, Rev YT
Example 3: Setting the Low Alarm Limit
Task Action
Set the analyzer to produce a “low
concentration alarm, when the moisture
concentration falls below 0 PPM.
Write one holding register (function 06).
Register address = 67 (holding register
#40068). Value = 0.
Example 4: Enabling “Hold During Verify”
Task Action
Enable hold the analog outputs at the last
measured value, when the analyzer is Oine
to perform a Verication.
Write one holding register (function 06).
Register address = 98 (holding register
#40099). Value = 1 (0-track during verify,
1-hold during verify).
Example 5: Setting the High-End of the Analog Output
Task Action
Set up the analog output so that a moisture
concentration of 100 PPM produces a 20 mA
current output.
Write one holding register (function 06).
Register address = 45 (holding register
#40046). Value = 100.
Example 6: Setting the Low-End of the Analog Output
Task Action
Set up the analog output so that a moisture
concentration of 1 PPM produces a 4 mA
current output.
Write one holding register (function 06).
Register address = 46 (holding register
#40047). Value = 100 (multiplied by scale of
100).
Example 7: Switching to “Sensor Saver” Mode
Task Action
Set the analyzer to operate in the Sensor
Saver mode.
Write one holding register (function 06).
Register address = 99 (holding register
#40100). Value = 1 (enable sensor saver =1,
disable sensor saver = 0).
6-12 | 3050-OLV Moisture Analyzer

Table of Contents

Other manuals for Ametek 3050-OLV

Related product manuals