PN 305200901, Rev YT
REG DEFINITION
UNITS SCALING
ACCESS
66
Concentration alarm range (Alarm1Hi= 2C Hex)
1
r,w
67
Concentration alarm range (Alarm1Lo= 2D Hex)
1
r,w
68
Concentration alarm range (Alarm2Hi= 2E Hex)
1
r,w
69
Concentration alarm range (Alarm2Lo= 2F Hex)
1
r,w
70
Process pressure units (PressUnits= 30 Hex)
1
r,w
71
Bypass valve state (BypassState= 31 Hex)
1
r,w
72
Zero coecient (FrequencyOset= 38 Hex)
Hz 1000 r,w
73
Gas based ow coecient (FlowCoe0= 39 Hex)
1000 r,w
74
Gas based ow coecient (FlowCoe1= 3A Hex)
1000 r,w
75
Process pressure upper limit (FlowCoe2= 3B Hex)
kPa
1
r,w
76
Flow meter parameter (FlowSpan= 3C Hex)
100 r,w
77
Auto verication hour (CalHour= 3D Hex)
1
r,w
78
Verication time in minutes (CalPeriod= 3E Hex)
minute
1
r,w
79
Auto verication day of month (CalMonthDays= 3F Hex)
1
r,w
80
Auto verication day of week (CalWeekDays= 40 Hex)
1
r,w
81
Hourly, daily, monthly auto verication (CalType= 41 Hex)
1
r,w
82
Software revision number (Version= 47 Hex)
100 r,w
83
Reset variable (SystemReset= 48 Hex)
1
r,w
84
Default memory ag (newFlag= 49 Hex)
1
r,w
85
Read only except test mode (AnalogInput= 4E Hex)
10 r,w
86
Read only except test mode (AnalogOutput= 4F Hex
1
r,w
87
Used in test mode (ReadTestFixture= 50 Hex)
1 r
88
Sample = 0,generator = 1,reference = 2 (ValveState= 56 Hex)
1
r,w
89
Current hour (Hour= 57 Hex)
hour
1
r,w
90
Current minute (Minute= 58 Hex)
minute
1
r,w
91
Current second (Second= 59 Hex)
second
1
r,w
92
Current month (Month= 5A Hex)
1
r,w
93
Current day (Day= 5B Hex)
1
r,w
94
Current year (Year= 5C Hex)
1
r,w
95
Used in test only (WriteTestFixture= 63 Hex)
1
96
Current day of week (WeekDay= 6D Hex)
1
r,w
97
Verication coecient (MoistureSpan= 6E Hex)
1000 r,w
98
Track or hold current output ag (HoldOut= 71 Hex)
1
r,w
99
Fast cycle =0, slow cycle=1 (SlowTiming= 72 Hex)
1
r,w
100
Minimum verication duration (VerDurationLimit= 73 Hex)
minute
1
r,w
101
Adjust span in verication ag (AdjustSpan= 74 Hex)
1
r,w
102
Alarm 1 enable ag (EnableAlarm1= 75 Hex)
1
r,w
103
Alarm 2 enable ag (EnableAlarm2= 76 Hex)
1
r,w
104 Reserved
Serial Communication Interface | 6-7