VNA Commands 5-2 :CALCulate Subsystem
S412E PM PN: 10580-00319 Rev. N 5-17
To set the trace 1 upper limit point to 5 Feet (trace 1 in distance domain
with current distance unit in meter):
:CALCulate:LIMit:UPPer:POINt:X 5 FT
OR to 4 Meter
:CALCulate1:LIMit:UPPer:POINt:X 4 M
:CALCulate:LIMit:UPPer:POINt:X 4
Related Command: :CALCulate:LIMit:TYPe
:CALCulate<Tr>:LIMit:UPPer:POINt:Y
Front Panel Access: Shift 6 (Limit), Limit Edit, Limit X
:CALCulate<Tr>:LIMit:UPPer:POINt:Y <y-parameter>
:CALCulate<Tr>:LIMit:UPPer:POINt:Y?
Description: Sets the location of the upper limit point of the given trace <Tr> on the
y-axis at the specified location. <Tr> is the trace number in the range
1 to 4. If no trace number is specified than defaults to trace number 1.
Sending the set command will change the Move Limit on the front panel
to Point if it is currently set to Limit and set the given trace as the
active trace. The <y-parameter> is defined in the given trace current
y-axis. If no unit is specified with the <y-parameter> then the default
unit will be used. The query version of the command returns the
location of the given trace upper limit point on the y-axis. If an error
occurs, such as limit not ON, the query version of the command returns
an error code of –400. Limit line must be on for the command to be
valid. Use :CALCulate:LIMit:TYPe to set the currently active limit
line.
Cmd Parameter: <NRf> <y-parameter> (depends on display type)
Query Response: <NR3> <y-parameter> (depends on display type)
Default Units: Current active trace y-axis unit
Related Command: :CALCulate:LIMit:TYPe
:CALCulate<Tr>:LIMit:UPPer:POINt:X
Front Panel Access: Shift 6 (Limit), Limit Edit, Amplitude