Spectrum Analyzer Commands 4-4 :DISPlay Subsystem
S412E PM PN: 10580-00319 Rev. N 4-23
4-4 :DISPlay Subsystem
This subsystem provides commands that modify the display of data for the user. They do not
modify the way in which data are returned to the controller.
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision <rel ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:PDIVision?
Title: Scale
Description: Sets the scale (dB/division) for the y-axis.
Parameter: <rel ampl>
Default Value: 10 dB/div
Default Unit: dB
Range: 1 dB to 15 dB
Front Panel Access: Amplitude, Scale
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel <amplitude>
:DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel?
Title: Reference Level
Description: Sets the reference level amplitude value for the y-axis. Note that this
may cause a change in attenuation if the automatic input attenuation
coupling is enabled.
Parameter: <amplitude>
Default Value: 10 dBm
Default Unit: Current active amplitude unit
Range: With reference level offset = 0 dB: 30 dBm to -150 dBm
Example: To set the reference level to 15 dBm (If y-axis is dBm)
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel 15
:DISPlay:WINDow:TRACe:Y:SCALe:RLEVel 15dBm
Related Command: :DISPlay:WINDow:TRACe:Y[:SCALe]:RLEVel:OFFset
Front Panel Access: Amplitude, Reference Level