4-3 :CONFigure Subsystem Spectrum Analyzer Commands
4-22 PN: 10580-00319 Rev. N S412E PM
:CONFigure:FSTRength
Title: Configure Field Strength
Description: Configures the default field strength measurement. Disables any other
active one-button measurements, including channel power, adjacent
channel power, occupied bandwidth, AM/FM demodulation and C/I.
Sets the antenna to the first antenna in the instrument’s antenna list.
Sets the instrument to single sweep mode (:INITiate:CONTinuous
OFF). Measurement settings can be modified by using the
[:SENSe]:FSTRength commands before initiating a sweep. Note that
this measurement is not valid in zero span.
Related Command: :FSTRength:ANTenna
:CONFigure:OBWidth
Title: Configure Occupied Bandwidth
Description: Configures the default occupied bandwidth measurement. Disables any
other active one-button measurements, including channel power,
ACPR, AM/FM demodulation and C/I. Sets the method to %. Sets the %
of power to 99%. Sets the instrument to single sweep mode
(:INITiate:CONTinuous OFF). Measurement settings can be modified
by using the [:SENSe]:OBWidth commands before initiating a sweep.
Note that this measurement is not valid in zero span.
Related Command: :OBWidth:STATe
:OBWidth:METHod
:OBWidth:PERCent
:OBWidth:XDB