VNA Commands 5-11 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 5-71
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
SHORt <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:SHORt?
Description: Sets the DUT capacitance value for the specified user.
Cmd Parameter: <NRf> <length> (millimeters)
Query Response: <NR3> <length> (millimeters)
Example: To set the DUT short offset for User 1 with line type coax to 3 mm:
:SENSe:CORRection:CKIT:USER:COAX:SOLT:SHORt 3
Front Panel Access: NA
[:SENSe]:CORRection:COLLect Subsystem
This subsystem controls the system calibration. To properly perform a calibration, several
parameters must be set. The table below lists all of the required commands. First, use the
:MEDium and :CONNector subcommands to specify the calibration line type and the DUT
port setup. Then use the :METHod and :TYPE subcommands to specify the calibration
method and the calibration type. Then use the :ACQUire subcommand to specify the
calibration components to be measured. Finally, use the :SAVe subcommand to calculate,
save, and finish the calibration. Note that the calibration components do not need to be
measured in any specific order.
[:SENSe]:CORRection:COLLect:ABORt:ALL
Description: Aborts the calibration measurement and restarts the current sweep or
measurement, or both.
Cmd Parameter: NA
Query Response: NA (no query)
Front Panel Access: NA