VNA Commands 5-2 :CALCulate Subsystem
S412E PM PN: 10580-00319 Rev. N 5-29
:CALCulate:MATH Subsystem
This subsystem contains functions for controlling math operations on the currently selected
measurement and memory.Trace Math Function
:CALCulate:MATH:FUNCtion NORMal|ADD|SUBTract|MULTiply|DIVide
:CALCulate:MATH:FUNCtion?
Description: Sets the math operations on the currently active trace and the trace
stored in memory. Note that a trace MUST be stored in Memory.
Setting the FUNCtion to NORMal is equivalent of setting the Trace
Math to “None” on the front panel. Setting the FUNCtion to ADD is
equivalent of setting the Trace Math to “Trace Plus Memory” on the
front panel. Setting the FUNCtion to SUBTract is equivalent to setting
the Trace Math to “Trace Minus Memory” on the front panel. Setting
the FUNCtion to MULTiply is equivalent to setting the Trace Math to
“Trace Multiply Memory” on the front panel. Setting the FUNCtion to
DIVide is equivalent to setting the Trace Math to “Trace Divide
Memory” on the front panel. The query version of the command returns
the string “NORM” for no trace math, “ADD” for trace plus memory,
“SUBT” for trace minus memory, “MULT” for trace multiply memory,
and “DIV” for trace divide memory.
Cmd Parameter: <char> NORMal|ADD|SUBTract|MULTiply|DIVide
Query Response: <char> NORM|ADD|SUBT|MULT|DIV
Default Value: NORM
Related Command: :CALCulate:MATH:MEMorize
Front Panel Access: Shift 5 (Trace), Trace Math
:CALCulate:MATH:MEMorize
Description: Copies the current measurement trace into memory.
Cmd Parameter: NA
Query Response: NA (no query)
Front Panel Access: Shift 5 (Trace), SaveTracetoMemory