Anritsu LMR Master S412E - [:Sense]:Correction:ckit:user Subsystem

Anritsu LMR Master S412E
704 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-11 [:SENSe] Subsystem VNA Commands
5-70 PN: 10580-00319 Rev. N S412E PM
[:SENSe]:CORRection:CKIT:USER Subsystem
This subsystem contains commands to configure the user device under test (DUT).
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
C[0]|1|2|3 <capacitance>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
C[0]|1|2|3?
Description: Sets the DUT capacitance value for the specified user.
Cmd Parameter: <NRf> <capacitance> (e-15, e-27, e-36, e-45)
Query Response: <NR3> <capacitance> (e-15, e-27, e-36, e-45)
Example: To set the DUT capacitance #1 for User 1 with line type coax to 5:
:SENS:CORR:CKIT:USER:COAX:SOLT:C1 5
Front Panel Access: NA
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
NAME <string>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:NAME?
Description: Sets the DUT name for the specified user.
Cmd Parameter: <string> (no parameter data or units)
Query Response: <string> (no parameter data or units)
Example: To set the DUT name for User 1 with line type coax and calibration
method SOLT:
:SENS:CORR:CKIT:USER:COAX:SOLT:NAME “SOLT1”
Front Panel Access: NA
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:
OPEN <length>
[:SENSe]:CORRection:CKIT:USER[1]|2|3|4:COAX:SOLT:OPEN?
Description: Sets the DUT capacitance value for the specified user.
Cmd Parameter: <NRf> <length> (millimeters)
Query Response: <NR3> <length> (millimeters)
Example: To set the DUT open offset for User 1 with line type coax to 3 mm:
:SENS:CORR:CKIT:USER:COAX:SOLT:OPEN 3
Front Panel Access: NA