6-5 [:SENSe] Subsystem Vector Voltmeter Commands
6-8 PN: 10580-00319 Rev. N S412E PM
6-5 [:SENSe] Subsystem
The commands in this subsystem relate to device-specific parameters, not to signal-oriented
parameters.
[:SENSe]:VVM:CABLe:SELect 1|2|3|4|5|6|7|8|9|10|11|12
[:SENSe]:VVM:CABLe:SELect?
Description: Selects the VVM cable. The query format of the command returns the
current VVM cable number.
Cmd Parameter: <char> 1|2|3|4|5|6|7|8|9|10|11|12
Query Response: <char> 1|2|3|4|5|6|7|8|9|10|11|12
Default Value: 1
Example: To set the Cable to 6:
:SENSe:VVM:CABLe:SELect 6
Front Panel Access: Table, Select Cable
[:SENSe]:VVM:FORMat DB|VSWR|IMPedance
[:SENSe]:VVM:FORMat?
Description: Sets the VVM Return type measurement format. The query format of
the command returns the VVM Return type measurement format.
Cmd Parameter: <char> DB|VSWR|IMPedance
Query Response: <char> DB|VSWR|IMP
Default Value: DB
Example: To set the type to VSWR:
:SENSe:VVM:FORMat VSWR
Front Panel Access: CW, Return Meas Format
[:SENSe]:VVM:FREQuency:CW <freq>
[:SENSe]:VVM:FREQuency:CW?
Description: Sets the VVM CW frequency. The query format of the command returns
the CW frequency.
Cmd Parameter: <NRf> <freq> (hertz)
Query Response: <NR3> <freq> (hertz)
Default Value: 500 kHz
Default Units: Hz
Range: 500 kHz to Unit Maximum Frequency
Front Panel Access: CW/Table, CW Frequency