Vector Voltmeter Commands 6-5 [:SENSe] Subsystem
S412E PM PN: 10580-00319 Rev. N 6-9
[:SENSe]:VVM:MODE CW|TABLe
[:SENSe]:VVM:MODE?
Description: Sets the VVM measurement mode. The query format of the command
returns the VVM measurement mode.
Cmd Parameter: <char> CW|TABLe
Query Response: <char> CW|TABL
Default Value: CW
Example: To set the mode to Table:
:SENSe:VVM:MODE TABLe
Front Panel Access: CW: Hard Key 1
TABLe: Hard Key 2
[:SENSe]:VVM:REFerence:CLEar
Description: Clears the reference data for the current port and measurement type.
Cmd Parameter: NA
Query Response: NA (no query)
Default Value: No Reference
Example: To clear the Reference:
:SENSe:VVM:REFerence:CLEar
Front Panel Access: CW/Table, Clear Reference
[:SENSe]:VVM:REFerence:MEMorize
Description: Sets the reference data for the current port and measurement type.
Cmd Parameter: NA
Query Response: NA (no query)
Default Value: No Reference
Example: To set the new Reference:
:SENSe:VVM:REFerence:MEMorize
Front Panel Access: CW/Table, Save New Reference